Our QS5 DL is loaded with advanced features like a Material Condition Analyzer, realtime BSCAN Graphics and a Coupling Level Indicator. And it also includes a Data Logger to send data and graphics stored in the unit memory to a PC using DMQ DataCenter software.
Use a QS5 DL to measure unpainted metals and metal structures such as tubes, pipes, storage tanks, valves, pressurized recipients and naval hulls to name a few. And it can also be used to measure plastics, acrylics, ceramics, and glass among other materials.
Modes: Pulse-Echo (normal measurements) Range: 0.6 to 500mm / 0.025 to 19.999in Depending on transducer and material Resolution: Up to 100mm, 0.01mm; or up to 3.936in, 0.001in From 100mm, 0.1mm; or from 3.936in, 0.01in V-Path Correction: Automatic Flight Time & Flight Time Resolution: 1 to 100us at 0.1us Materials Velocity Range: 100 to 19.999 m/s 2.0 in/ms at 787.0 in/ms Velocity Resolution: 1 m/s – 0.1 in/ms Sampling Rate: Selectable 4, 8, or 20Hz (Fast) Emission Pulse: Adjustable 20 to 220V
Metric: Millimeters (mm) Imperial: Inches (in) Temporary: Microseconds (us)
Coupling: Coupling level indicator Material Condition Analyzer: Alerts on possible corrosion or pitting Selectable Scanning Speeds: 4, 8, 20Hz (Fast) BSCAN Graphic Representations: In real time Gain and Trigger: Adjustable Capture Minimum: Hold minimum reading Differential Mode: Difference between actual measurement and reference value Percentage Differential Mode: Measured value represented as a percentage of the reference value Erase Mode: Erases measurement when transducer is lifted View Modes: 3 numerical modes: different information is displayed Nums Zoom mode: large numbers (17mm tall) Graphic mode: view real time B-Scan graphics Internal Clock: Date and Time (Selectable format)
Types: Maximum, Minimum, Range, Average Error: High, Low, High%, Low% Distribution: Values represented in a graphic format
Calculations are based on stored values only
Low & High: User defined Material Condition Analyzer: Alerts for possible presence of corrosion or pitting Alarm Types: Visible: flashing backlight or irregular digits Audible: intermitent beep
Digits: 17mm / 0.66in tall Graphic LCD: 128×64 pixels Viewable Area: 45 x 30mm / 1.77 x 1.18in Backlight: LED backlight (ON, AUTO, OFF) Contrast: Digital 30 steps
Languages: English, Spanish, Portuguese Menu Screens: Icon based User ID: Personal information with password protection
Type: Dual Crystal Connectors: Type Lemo 00 Frequencies: 2, 5, 7MHz
1 Point: Auto with no Cal key using the unit test block (6.35 +/- 0.01mm or 0.25in +/- 0.001in) 2 Point: Use the unit test block and 2 calibration blocks Materials Velocity Chart: On unit screen
Upgrade to a QS5 E or DLE with the purchase of a model upgrade license and a hi-damp transducer
Storage: Up to 32000 readings Data Transfer: Send to a PC using DMQ DataCenter software included at no additional charge Organization: 8 files with 10 character alphanumeric naming Capture Modes: Manual: Touch the “Store” key to store values Continuous: Automatically stores all values picked up by the transducer Each: Automatically stores single values each time the transducer is coupled Date and Time Register: Register day and time
Connectivity: USB (cable included) or RS232 (optional cable)
Touch-Sense Keypad: No mechanical parts Keypad: Sensitivity adjustment Auto pressure detection Auto scrolling detection
Dimensions: H: 117mm / 4.60in L: 78.74mm / 3.10in W: 24mm / 0.94in Weight: 200g / 7.05oz with batteries Working Temperature: -10°C to 50°C / 14°F to 122°F Storage Temperature: -20°C to 60°C / -4°F to 140°F Enclosure: High impact ABS with rubber sides for easy grip Front Panel: Polycarbonate
Batteries: 2 x AA 1.5v Alkaline 2 x AA 1.2v Ni-MH Rechargeable Battery Life: 100 hours with backlight off Battery Indicator: On all screens with a low level alarm Auto-Off: Programmable 1 to 120 minutes
Electronic Unit: 5 year limited warranty Transducers: 6 months